FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2072304 · Received April 28, 2011

Report

Report Number
2024168-2011-02989
Event Type
Injury
Date Received
April 28, 2011
Date of Event
April 5, 2011
Report Date
April 6, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). TWO UNUSED REPRESENTATIVE SAMPLES WITH THE SAME PART AND LOT NUMBER WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED ON BOTH OF THE RETURNED DEVICES AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. A ROOT CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE RETURN OF THE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED EVENT. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT. IN ADDITION, A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND THERE HAVE BEEN NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, FOLLOWING NEEDLE DEPLOYMENT, WHEN THE PLUNGER WAS PULLED BACK THE WHOLE DEVICE WAS PULLED FROM THE ANATOMY, WITHOUT CAPTURING THE SUTURES. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 020226H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention