FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2072299
·
Received April 28, 2011
Report
- Report Number
- 2050012-2011-01257
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 31, 2011
- Report Date
- March 31, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(4) 2011 AND FOUND THE 10PSI LINE CONNECTED TO FITTING ON NO FOAM CAP WAS LOOSE AND SECURED TUBING ON FITTING. FSE DID NOT SEE FURTHER EVIDENCE OF LEAKING THROUGH THE DAY SHIFT. NO LOOSE OR CRACKED FITTINGS WERE FOUND IN THE HYDRO DRAWER.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT A PUDDLE HAD FORMED UNDER THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE PUDDLE EVAPORATED RAPIDLY AND LEFT NO RESIDUE. CUSTOMER STATED THE TUBING AND FITTING FOR THE NO FOAM CAP WAS LOOSE. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |