FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2072299 · Received April 28, 2011

Report

Report Number
2050012-2011-01257
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 31, 2011
Report Date
March 31, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(4) 2011 AND FOUND THE 10PSI LINE CONNECTED TO FITTING ON NO FOAM CAP WAS LOOSE AND SECURED TUBING ON FITTING. FSE DID NOT SEE FURTHER EVIDENCE OF LEAKING THROUGH THE DAY SHIFT. NO LOOSE OR CRACKED FITTINGS WERE FOUND IN THE HYDRO DRAWER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT A PUDDLE HAD FORMED UNDER THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE PUDDLE EVAPORATED RAPIDLY AND LEFT NO RESIDUE. CUSTOMER STATED THE TUBING AND FITTING FOR THE NO FOAM CAP WAS LOOSE. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO NA