ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-02983
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- April 4, 2011
- Report Date
- April 4, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. ALL KEYPAD BUTTONS WERE RESPONDING INTERMITTENTLY. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE CONTACTS. MANUFACTURE DATE: 09/07/2007.
ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT DENIED THAT THE PUMP WAS EXPOSED TO MOISTURE. HE CLAIMED THAT THE KEYPAD WAS INTACT AND THE BUTTONS WERE SPRINGING BACK AS USUAL. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS NOT RESPONDING APPROPRIATELY TO KEYPAD BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 |