FDA Adverse Event Death Summary report: N

QUICK FLEX LV LEAD

MDR report key: 2072255 · Received April 28, 2011

Report

Report Number
2017865-2011-02494
Event Type
Death
Date Received
April 28, 2011
Date of Event
February 17, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
OJX
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPIRED ON 02/17/2011. THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK FLEX LV LEAD PERMANENT PACEMAKER ELECTRODE OJX ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1156T/86 NA

Patients

Seq Age Sex Outcome Treatment
1 84 YR Death 3207-36, 547770, 7120, AHA31401, 1699TC, EP033569