FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 2072214 · Received April 28, 2011

Report

Report Number
2031642-2011-00135
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 14, 2011
Report Date
April 1, 2011
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

NO SERVICE REQUIRED; UNIT OUT OF WARRANTY. USER ERROR; CORRECTIVE ACTION TAKEN BY CUSTOMER. NO SERVICE REQUIRD; UNIT OUT OF WARRANTY.

Description of Event or Problem · 1

THE HOSPITAL BIOMEDICAL ENGINEER REPORTED THE VENTILATOR AC POWER CORD WAS NOT PLUGGED INTO THE AC OUTLET. THE BIOMEDICAL ENGINEER REPORTED THE TECHNICIAN HAD PLUGGED THE HUMIDIFIER INTO THE AC OUTLET INSTEAD OF THE VENTILATOR. AS A RESULT, THE VENTILATOR WAS BEING POWERED VIA THE BACKUP BATTERY. THE VENTILATOR OPERATED ON BACKUP BATTERY UNTIL THE BACKUP BATTERY DEPLETED, AND THE VENTILATOR THEN SHUT DOWN AND ALARMED. REVIEW OF THE VENTILATOR'S DIAGNOSTIC LOG VERIFIED THE DEVICE WAS OPERATING ON BACKUP BATTERY POWER AND THE BACKUP BATTERY FUNCTIONED UNTIL IT DEPLETED, AT WHICH TIME THE VENTILATOR WILL SHUT DOWN AND ALARM AS SPECIFIED DUE TO LOSS OF POWER. THERE WAS NO PATIENT HARM REPORTED, AND THE TECHNICIAN WAS AT THE BEDSIDE AND PLACED THE PATIENT ON ANOTHER VENTILATOR. THIS EVENT IS BEING REPORTED AS A USER ERROR. PER THE VENTILATOR'S OPERATORS MANUAL, WHEN THE VENTILATOR IS POWERED BY BACKUP BATTERY IT WILL GENERATE A NONRESETABLE, NONSILEANCEABLE ALARM EVERY 60 SECONDS. DURING THIS STATE A BATTERY IN USE YELLOW LED IS LIT. OPERATION WILL CONTINUE UNTIL THE BATTERY HAS 5 MINUTES OF OPERATION LEFT. A RED BATTERY LOW LED WILL FLASH AND A NONRESETABLE HIGH PRIORITY ALARM WILL SOUND. WHEN THE BATTERY LOW INDICATOR IS FLASHING RED, OPERATION OF THE VENTILATOR FROM THE BATTERY POWER SHOULD BE DISCONTINUED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1