FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2072210 · Received April 28, 2011

Report

Report Number
2531779-2011-02971
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 2, 2011
Report Date
April 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER THE EVALUATION IS COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 12/20/2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED THAT REBOOTING HAD OCCURRED. NO PHYSICAL DAMAGE WAS NOTED TO THE PUMP. THE PUMP BOOTED TO THE VERIFY SCREEN WITH APPROPRIATE ALARMS. THE BATTERY CAP WAS TESTED; REBOOTING WAS DUPLICATED AND THE BATTERY CAP CONTACTS WERE FOUND TO BE OUT OF SPECIFICATION. A TEST BATTERY CAP WAS INSERTED TO CONTINUE TESTING. THE PUMP CURRENT DRAWS WERE TESTED AND FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT REBOOTING. THE PUMP WAS OPENED AND AN ELECTRICAL COMPONENT WAS FOUND TO HAVE BROKEN OFF THE PCB.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED SELF-REBOOT ISSUES. THERE WAS NO REPORT OF PRODUCT MISUSE. THE REPORTER INDICATED THERE WAS NO CORROSION OR MOISTURE IN THE BATTERY COMPARTMENT, THE BATTERY CAP IS INTACT, THERE WAS NO STRUCTURAL DAMAGE, AND THERE WAS SEVERAL BATTERY ALARM MESSAGE SINCE THE ANIMAS PUMP WAS INITIALLY DROPPED. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR