ARCHITECT I1000SR ANALYZER
Report
- Report Number
- 1628664-2011-00210
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Report Date
- October 21, 2009
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). A PRODUCT DEFICIENCY HAS BEEN IDENTIFIED. WITH A UNIQUE COMBINATION OF INSTRUMENT AND REAGENT CONDITIONS THERE IS A POTENTIAL FOR SAMPLE CARRYOVER WHICH CAN RESULT IN FALSELY ELEVATED B-HCG CONCENTRATIONS ON THE ARCHITECT I1000SR SYSTEM. NEGATIVE SAMPLES HAVE RETURNED BOTH POSITIVE AND GREY ZONE B-HCG RESULTS WHEN A HIGH CONCENTRATION B-HCG SAMPLE IS ASSESSED PRIOR TO THE NEGATIVE SAMPLE WITH ARCHITECT TOTAL B-HCG (LIST NUMBER 7K78/6C21) ON AN I1000SR INSTRUMENT WHICH ALSO UTILIZES THE ARCHITECT RUBELLA IGG (LIST NUMBER 6C17) ASSAY. FURTHER INVESTIGATION OF THIS QUALITY ISSUE WILL BE CONDUCTED. AN INVESTIGATION IS IN PROCESS.
THE CUSTOMER STATED AN ARCHITECT I1000SR ANALYZER GENERATED FALSELY ELEVATED B-HCG RESULTS FOR FIVE PATIENT SAMPLES. DATA WAS REVIEWED AND FOUND NOT TO AFFECT THE CLINICAL INTERPRETATION OR MEDICAL DECISION MAKING. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I1000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARCH I1000SR LN 01L86-01 SN (B)(6) | ARCHITECT TOTAL B-HCG REAGENT, LN 7K78-25| ARCHITECT TOTAL B-HCG REAGENT, LN 7K78-25 |