FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2072080 · Received April 21, 2011

Report

Report Number
1824206-2011-02327
Event Type
Malfunction
Date Received
April 21, 2011
Date of Event
March 23, 2011
Report Date
March 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A NEW SIDERAIL WAS SENT TO THE ACCOUNT TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT'S MAINTENANCE ALLEGED THE LEFT HEAD RAIL WAS TORN OFF OF THE BED. UNK CAUSE. THE ACCOUNT'S MAINTENANCE STATED HE FOUND THE BED THIS WAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1