FDA Adverse Event
Malfunction
Summary report: N
ADVANTA BED
MDR report key: 2072057
·
Received April 21, 2011
Report
- Report Number
- 1824206-2011-02313
- Event Type
- Malfunction
- Date Received
- April 21, 2011
- Date of Event
- March 29, 2011
- Report Date
- March 29, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN INDICATED THAT IT LOOKED LIKE THE CABLE WAS PINCHED BETWEEN THE MOVING PARTS OF THE HEADRAIL. HE REPLACED BOTH SIDERAIL CABLE ASSEMBLIES TO REPAIR THE BED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE RIGHT HEAD SIDERAIL CABLE IS CUT/FRAYED AND METAL WIRING IS EXPOSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |