FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2072057 · Received April 21, 2011

Report

Report Number
1824206-2011-02313
Event Type
Malfunction
Date Received
April 21, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN INDICATED THAT IT LOOKED LIKE THE CABLE WAS PINCHED BETWEEN THE MOVING PARTS OF THE HEADRAIL. HE REPLACED BOTH SIDERAIL CABLE ASSEMBLIES TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE RIGHT HEAD SIDERAIL CABLE IS CUT/FRAYED AND METAL WIRING IS EXPOSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1