COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2011-03251
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 22, 2011
- Report Date
- March 25, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. ADDITIONAL: THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4).
BAXTER FIELD SERVICE TECHNICIAN SERVICED A COLLEAGUE INFUSION PUMP FOR A DAMAGED BATTERY. THE PROCESS STEP WAS UNKNOWN AND ANY REPORT OF PATIENT INVOLVEMENT IS UNKNOWN. THERE IS NO PATIENT INJURY REPORTED. THE BAXTER FIELD SERVICE TECHNICIAN REPAIRED THE DEVICE ON SITE. AS SUCH, THE DEVICE WILL NOT BE RETURNED TO (B)(4) TECHNICAL SERVICES. DURING THE REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.08.92, CATEGORIZED AS REMEDIATED.