GATEWAY BALLOON CATHETER
Report
- Report Number
- 2939204-2011-00261
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- April 21, 2011
- Report Date
- April 21, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- GBA
- PMA / PMN Number
- H050001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
A DEVICE HISTORY RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. INSPECTION OF THE RETURNED DEVICE UNDER MAGNIFICATION FOUND THAT THE INNER SHAFT WAS CORRECTLY SEATED IN THE HUB AND ADHESIVE WAS PRESENT IN THE ADHESIVE CAVITY. IT WAS NOTED THAT THERE WAS A SMALL SLIT IN THE INNER SHAFT THAT DIRECTLY ALIGNED WITH THE INFLATION PORT. NO OTHER ANOMALIES WERE NOTED. THE DAMAGE TO THE INNER SHAFT MAY BE DUE TO PERFORATION WITH A NEEDLE OR GUIDEWIRE USED DURING THE PROCEDURE. NO TOOLS ARE INSERTED IN THE MANIFOLD INFLATION PORT AND ALL DEVICES ARE TESTED FOR BALLOON INFLATION AND DEFLATION PERFORMANCE DURING MANUFACTURE. THEREFORE, IT WAS DETERMINED THAT THE REPORTED EVENT WAS DUE TO HANDLING OF THE DEVICE IN THE PROCEDURE.
DURING PREPARATION IT WAS FOUND THAT THE BALLOON LEAKED. THERE WAS NO PATIENT INVOLVEMENT.
DURING PREPARATION IT WAS FOUND THAT THE BALLOON LEAKED. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GATEWAY BALLOON CATHETER | CATHETER, BALLOON TYPE | GBA | BOSTON SCIENTIFIC - MAPLE GROVE | M0032072409200 | 3916230 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |