FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2071886
·
Received April 28, 2011
Report
- Report Number
- 2531779-2011-02957
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 30, 2011
- Report Date
- March 30, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE UP/DOWN KEYPAD BUTTONS HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |