FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2071784 · Received April 28, 2011

Report

Report Number
2531779-2011-02953
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MFR DATE 05/2007. 2531779-03/24/2010-003-R. PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR.

Description of Event or Problem · 1

PUMP WAS RETURNED TO ANIMAS FOR A SHORT LOAD STEP CAUSING A LARGE PRIME VOLUME. INVESTIGATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 51