FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 2071466 · Received April 28, 2011

Report

Report Number
2124215-2011-07454
Event Type
Injury
Date Received
April 28, 2011
Date of Event
January 19, 2011
Report Date
January 20, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

OUR RECORDS INDICATE THAT THIS LEAD REMAINS IMPLANTED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION INDICATED THAT A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SURGICALLY ABANDONED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD EXHIBITED LOSS OF CAPTURE, INCREASED THRESHOLD MEASUREMENTS, AND HIGH PACING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS. THE PHYSICIAN DECIDED TO CONTINUE MONITORING THE LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4087

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R 0158| 4513| H179| N118| 4087| 4518