FDA Adverse Event Malfunction Summary report: N

LIGAMAX-APPLIER

MDR report key: 2071453 · Received April 28, 2011

Report

Report Number
3005075853-2011-01743
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 5, 2011
Report Date
April 8, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DATE SENT: (B)(4) 2011. FILE IS NOT REPORTABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE WAS USED DURING AN UNKNOWN PROCEDURE. AFTER THREE CLIPS DELIVERED PROPERLY, THE DEVICE STUCK, IT WAS IMPOSSIBLE TO ACTIVATE IT. HANDLES WOULD NOT CLOSE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Description of Event or Problem · 1

NEW INFORMATION ADVISING THAT "DEVICE WAS STUCK" INDICATING THE HANDLES WOULD NOT CLOSE WHICH REFLECTS A FIRING ISSUE. BASED ON NEW INFORMATION, FILE IS NOT REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK G4U05A

Patients

Seq Age Sex Outcome Treatment
1