FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2071309 · Received April 27, 2011

Report

Report Number
2050012-2011-01324
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 27, 2011
Report Date
March 28, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER THE CUSTOMER, SAMPLE 1 NOTICED TO HAVE CLOTS IN IT WHILE REFRIGERATED AFTER SAMPLE RUN. SAMPLE 2 WAS NORMAL LOOKING AFTER SAMPLE WAS RUN. NO MENTION OF CALIBRATION OR QC PROBLEMS BY THE CUSTOMER. THE CUSTOMER CLEANED THE GLUCOSE CUP ASSEMBLY AND IT APPEARED (ABNORMALLY) YELLOW IN COLOR. ELECTRODE FACE WAS CLEANED. MONTHLY MAINTENANCE IS PERFORMED WEEKLY BY THE CUSTOMER. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE GLUCOSE MODULE. THE MODULE HAS BEEN RETURNED TO BCI FACILITY FOR FURTHER EVALUATION. ROOT CAUSE REMAINS UNKNOWN AT THIS TIME.

Description of Event or Problem · 1

A GLUCOSE POST-MOD CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI), REPORTING TWO (2) ERRONEOUSLY HIGH GLUCOSE (GLUCM) PATIENT RESULTS THAT WERE GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER WHILE RUNNING IN DUPLICATE MODE. THE GLUCOSE POST-MOD ACCOUNT IS MONITORED THROUGH (B)(4). PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT AS THE CUSTOMER RUNS GLUCOSE SAMPLES IN DUPLICATE MODE AND VERIFIES RESULTS PRIOR TO REPORTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1