FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2071251 · Received April 27, 2011

Report

Report Number
1423500-2011-05097
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
September 30, 2009
Report Date
April 7, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE IIPV IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW/ NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER; HOWEVER, THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN ISSUE OF AN INCREASED INTRAPERITONEAL VOLUME (IIPV) EVENT WAS FOUND IN THE PATIENT EVENT LOGS: OCCURRENCE DATE (B)(6) 2009 WITH DRAIN VOLUME OF 3611 MILLILITERS (ML) DURING CYCLE 4. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1