FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2071250 · Received April 27, 2011

Report

Report Number
2531779-2011-02944
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 28, 2011
Report Date
March 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER THE EVALUATION IS COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 06/21/2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/23/2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX HISTORY CONFIRMED MULTIPLE POWER LOSS EVENTS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT SELF-REBOOTING OCCURING.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO AN ALLEGED SELF-REBOOT ISSUE. THE ANIMAS INSULIN PUMP ALLEGEDLY REBOOTED DAILY. THERE WAS NO REPORT OF PRODUCT MISUSE. THE BATTERY CAP WAS REPLACED BUT THE ISSUE PERSISTED. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR