PRESSURE MONITORING KIT
Report
- Report Number
- 2015691-2011-15349
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- March 29, 2011
- Report Date
- March 29, 2011
- Manufacturer
- EDWARDS LIFESCIENCES DR
- Product Code
- DXO
- PMA / PMN Number
- K925638
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION INDICATED THAT THE ALTHOUGH THE ESTIMATED BLOOD LOSS WAS NOT QUANTIFIABLE, THERE WAS NO PATIENT INJURY AS A RESULT. THE ALARM ON THE MONITOR RANG WHEN THE ARTERIAL LINE DISAPPEARED AND THIS OCCURRED "OVER A MINUTE OR SO IT IS THOUGHT".
RECEIVED (1) INCOMPLETE VAMP ADULT SYSTEM. DRY BLOOD WAS VISIBLE THROUGHOUT THE SYSTEM. THE ACTUAL COMPLAINT WAS NOT CONFIRMED; THE TUBING DID NOT BREAK BUT WAS COMPLETELY DETACHED FROM THE SOLVENT BOND JOINT WITH THE Z-SITE. THE DETACHED TUBING WAS NOT RETURNED WITH THE VAMP SYSTEM. INDICATION OF BONDING SOLVENT WAS VISIBLE AT SMALL PART OF Z-SITE BOND AREA. THE INNER DIAMETER OF THE Z-SITE TUBE HOUSING (BOND AREA) WAS MEASURED AND FOUND WITHIN THE ALLOWABLE SPECIFICATION. IT WAS NOT POSSIBLE TO MEASURE THE TUBING OUTER DIAMETER OR THE BONDING CONDITION OF THE DETACHED TUBING SINCE THE TUBING WAS NOT RETURNED WITH THE SYSTEM. VISUAL EXAMINATION WAS PERFORMED UNDER 10X AND 40X MAGNIFICATION. THE TUBING DETACHMENT OCCURRED ON THE PATIENT SIDE OF THE PRESSURE MONITORING KIT. THE REPORTED DEFECT WAS CONFIRMED TO BE RELATED TO THE MANUFACTURING PROCESS. ACTIONS WERE OPENED TO ADDRESS COMPLAINTS OF THIS TYPE.
IT WAS REPORTED THAT THE TUBING BROKE OFF BELOW THE Z SITE AND THE PATIENT STARTED BLEEDING. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRESSURE MONITORING KIT | DISPOSABLE PRESSURE TRANSDUCER | DXO | EDWARDS LIFESCIENCES DR | PXVMP120 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |