FDA Adverse Event Death Summary report: N

CAPSURE SP NOVUS

MDR report key: 2071211 · Received April 27, 2011

Report

Report Number
2649622-2011-07344
Event Type
Death
Date Received
April 27, 2011
Date of Event
August 3, 2000
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE DEVICE AND LEADS ARE IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED AND ANALYZED AND FOUND TO HAVE NORMAL BATTERY DEPLETION. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

AN IMPLANTABLE PULSE GENERATOR (IPG) AND TWO LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATION THE PATIENT DIED APPROXIMATELY TWO MONTHS POST THE IMPLANT OF THE IPG SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5092 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Death