FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2071158 · Received April 27, 2011

Report

Report Number
2531779-2011-02935
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 20, 2011
Report Date
March 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). FOR DATE OF EVENT, THE USER STATED THE ISSUE OCCURRED AROUND ONE MONTH PRIOR TO CALLING ANIMAS ((B)(6) 2011). THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED ADHESIVE UNDER KEYPAD BUTTON CONTACTS. ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED.

Description of Event or Problem · 1

THE PATIENT'S MOTHER REPORTED THAT THE UP, DOWN, AND OK BUTTON CONTACTS REQUIRE SEVERAL PRESSES TO ACHIEVE A PUMP RESPONSE. THE ISSUE REPORTEDLY STARTED A MONTH PRIOR TO CONTACTING ANIMAS AND GOT PROGRESSIVELY WORSE. THE KEYPAD RUBBER IS REPORTEDLY INTACT AND THE USER DENIED ANY EXPOSURE TO CLEANING PRODUCTS. THE BUTTONS REPORTEDLY FELT FLAT AND DO NOT BOUNCE OR SPRING BACK NORMALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 13 YR