FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2071158
·
Received April 27, 2011
Report
- Report Number
- 2531779-2011-02935
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- March 20, 2011
- Report Date
- March 29, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(6). FOR DATE OF EVENT, THE USER STATED THE ISSUE OCCURRED AROUND ONE MONTH PRIOR TO CALLING ANIMAS ((B)(6) 2011). THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED ADHESIVE UNDER KEYPAD BUTTON CONTACTS. ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED.
Description of Event or Problem · 1
THE PATIENT'S MOTHER REPORTED THAT THE UP, DOWN, AND OK BUTTON CONTACTS REQUIRE SEVERAL PRESSES TO ACHIEVE A PUMP RESPONSE. THE ISSUE REPORTEDLY STARTED A MONTH PRIOR TO CONTACTING ANIMAS AND GOT PROGRESSIVELY WORSE. THE KEYPAD RUBBER IS REPORTEDLY INTACT AND THE USER DENIED ANY EXPOSURE TO CLEANING PRODUCTS. THE BUTTONS REPORTEDLY FELT FLAT AND DO NOT BOUNCE OR SPRING BACK NORMALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | OTP GLUCOSE MGMT SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |