ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-02936
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- March 30, 2011
- Report Date
- March 30, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED THAT ONE INSTANCE OF REBOOTING HAD OCCURRED, WHICH COULD NOT BE DUPLICATED ON INVESTIGATION. THERE WERE NO ALARMS RELATED TO THE COMPLAINT IN THE PUMP HISTORY. EVALUATION FOUND THAT THE PUMP POWERS UP PROPERLY, AND THERE WAS NO DAMAGE FOUND TO THE BATTERY CAP, BATTERY COMPARTMENT, OR TO THE POWER CIRCUIT. UNRELATED TO THE POWER LOSS ALLEGATION, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED.
THE REPORTER REPORTED, THE PUMP LOST POWER AND SELF-REBOOTED. THE PATIENT EXPERIENCED NO SYMPTOMS OF HIGH OR LOW BLOOD GLUCOSE LEVELS DUE TO THIS PUMP ISSUE. TROUBLESHOOTING REVEALED THERE WERE NO CRACKS IN THE BATTERY COMPARTMENT, NO LEAKS OR CORROSION, AND THE BATTERY CAP WAS INTACT AND SECURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | OTP GLUCOSE MGMT SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |