FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2071143 · Received April 27, 2011

Report

Report Number
2531779-2011-02932
Event Type
Malfunction
Date Received
April 27, 2011
Report Date
March 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED CONTAMINATION UNDER ALL KEYPAD BUTTON CONTACTS. ALL KEYPAD BUTTONS INTERMITTENTLY ACTIVATED PUMP FUNCTIONS WHEN PRESSED.

Description of Event or Problem · 1

THE PATIENT'S MOTHER REPORTED THAT FOR SIX MONTHS THE UP AND DOWN ARROW BUTTONS HAVE BEEN SLOW TO ACTIVATE PUMP FUNCTIONS WHEN PRESSED. SHE REPORTS THAT THE BUTTONS WILL NO LONGER ACTIVATE PUMP FUNCTIONS AT ALL. SHE REPORTED THAT THE KEYPAD RUBBER WAS INTACT AND THE BUTTONS CLICKED AND SPRUNG BACK NORMALLY. SHE DENIED ANY EXPOSURE TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 11 YR