FDA Adverse Event Death Summary report: N

ENTRUST AT

MDR report key: 2071142 · Received April 27, 2011

Report

Report Number
6000144-2011-02053
Event Type
Death
Date Received
April 27, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEADS IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY - (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS RESULTS REVEALED NO ANOMALIES FOUND. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THE OUTER INSULATION HAD A COSMETIC CUT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEADS IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY - (B)(4) NO ANOMALIES FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY - (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS RESULTS REVEALED NO ANOMALIES FOUND. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THE OUTER INSULATION HAD A COSMETIC CUT. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

ASKU

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD FRACTURED. THE LEAD INTEGRITY ALERT WAS LOADED AND THERE WERE FOUR INAPPROPRIATE SHOCKS. THE LEAD WAS NOT GOING TO BE REPLACED AS THE PATIENT WAS ON HOSPICE. IT WAS REPORTED THE PATIENT HAD DIED WITH CAUSE OF DEATH REPORTED TO BE HYPERTENSION, DIABETES, AND CONGESTIVE HEART FAILURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND NOT RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD FRACTURED. THE LEAD INTEGRITY ALERT WAS LOADED AND THERE WERE FOUR INAPPROPRIATE SHOCKS. THE LEAD WAS NOT GOING TO BE REPLACED AS THE PATIENT WAS ON HOSPICE. IT WAS REPORTED THE PATIENT HAD DIED WITH CAUSE OF DEATH REPORTED TO BE HYPERTENSION, DIABETES, AND CONGESTIVE HEART FAILURE. FOLLOW UP WITH THE CLINIC REPORTED THE HIGH POWER LEAD HAD BEEN DEACTIVATED TWO MONTHS PRIOR TO THE PATIENT'S DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Death| O