FDA Adverse Event Malfunction Summary report: N

OTHER

MDR report key: 2071071 · Received April 27, 2011

Report

Report Number
2531779-2011-02929
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 06/07/2011 - DEVICE EVALUATION: THE CARTRIDGES HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/07/2011 WITH THE FOLLOWING FINDINGS: CARTRIDGES WERE RETURNED FROM LOT#'S B201617 & B201627. THE CARTRIDGES PASSED VISUAL INSPECTION, NO DAMAGE OR DEFECTS WERE NOTED TO THE LUER CONNECTIONS OR O-RINGS. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTIONS, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGES.

Additional Manufacturer Narrative · 1

BRAND NAME: CARTRIDGE. LOT #:B201627. THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER REPORTED THE PATIENT NOTED THE INSULIN CARTRIDGE WAS LEAKING AND THE PUMP AND CARTRIDGE WERE WET. ON (B)(6) 2011 THE PATIENT OBTAINED BLOOD GLUCOSE READINGS THAT WERE HIGHER THAN EXPECTED, SUCH AS 254 MG/DL. THE PATIENT EXPERIENCED THE SYMPTOMS OF FEELING TIRED AND "OUT OF IT"; SHE TESTED NEGATIVE FOR KETONES AND EXPERIENCED NO SYMPTOMS SUGGESTING SEVERE HYPERGLYCEMIA. THE PATIENT CONTACTED HER PHYSICIAN FOR ASSISTANCE/ADVICE, AND WAS ADVISED TO MONITOR HER BLOOD GLUCOSE LEVELS AND CONTACT ANIMAS; SHE RECEIVED NO TREATMENT. TROUBLESHOOTING REVEALED THERE WAS NO DAMAGE TO THE CARTRIDGE OR TUBING AND NO CRACKS IN THE PUMP CASING; THE TUBING WAS SECURE TO THE CARTRIDGE LUER LOCK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OTHER INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1 26 YR