FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 2071037 · Received April 27, 2011

Report

Report Number
6000001-2011-03235
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 10, 2011
Report Date
March 11, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE COULD NOT BE DETERMINED AT THIS TIME. NO REPAIRS HAVE BEEN PERFORMED AS THE REFERENCED DEVICE HAS BEEN REMOVED FROM SERVICE. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED THAT THE DEVICE HAS NOT BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH CAPA (B)(4).

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH FAILURE CODE 810:11 OCCURRED ON CHANNEL A. THE EVENT OCCURRED UPON POWER ON IN THE INTENSIVE CARE UNIT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING THE PRODUCT EVALUATION AT BAXTER, IT WAS DISCOVERED THAT FAILURE CODE 810:11 OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.04.00, CATEGORIZED AS UNREMEDIATED.