JAPANESE BASAL/BOLUS INFUSORLV 4X4 WITH 30 MIN LOCKOUT
Report
- Report Number
- 6000001-2011-03230
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- April 13, 2011
- Report Date
- April 14, 2011
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). NO FLOW CONDITION NOT CONFIRMED. VISUAL EXAMINATION OF THE UNIT FOUND NO SIGNS OF ABNORMALITY ON THE BLADDER OR ON OTHER PARTS. A FUNCTIONAL TEST WAS SUBSEQUENTLY PERFORMED BY MANUALLY FILLING THE BLADDER WITH DEXTROSE SOLUTION. NO SIGNS OF EXCESSIVE AIR WERE NOTED IN THE BLADDER DURING AND AFTER FILL. FLOW WAS READILY OBSERVED AT THE LUER AFTER FILL. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT.
(B)(4). ADDITIONAL NARRATIVE:THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
BAXTER (B)(4) RECEIVED A REPORT THAT ONE (1) (B)(4) BASAL/BOLUS INFUSORLV 4X4 WITH 30 MIN LOCKOUT DEVICE WAS OBSERVED WITH EXCESS AIR IN THE BLADDER DURING FILLING. THE DEVICE WAS BEING FILLED WITH 300ML OF ROPIVACAINE HYDROCHLORIDE HYDRATE AND 6ML OF FENTANYL CITRATE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | JAPANESE BASAL/BOLUS INFUSORLV 4X4 WITH 30 MIN LOCKOUT | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10J067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FENTANYL CITRATE| ROPIVACAINE HYDROCHLORIDE HYDRATE |