FDA Adverse Event Malfunction Summary report: N

ENDOPATH** ILS ENDOSCOPIC CURVED INTRALUMINAL STAPLER - 29 MM

MDR report key: 2070970 · Received April 27, 2011

Report

Report Number
3005075853-2011-01750
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
April 8, 2011
Report Date
April 11, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
KOG
PMA / PMN Number
K940967
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4): UNCUT WASHER - BLEMISHED. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE, THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION, IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (PLASTIC TO PLASTIC), STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. FOR MORE INFORMATION PLEASE REFER TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT FIRE. NO OTHER DETAILS OF THE EVENT ARE KNOWN. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** ILS ENDOSCOPIC CURVED INTRALUMINAL STAPLER - 29 MM CIRCULAR STAPLERS KOG ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1