FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2070860 · Received April 15, 2011

Report

Report Number
2531779-2011-02594
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
February 1, 2011
Report Date
March 21, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE LETTERING ON THE KEYPAD HAS WORN OFF BUT THE RUBBER KEYPAD COVER WAS INTACT. ALL KEYPAD BUTTONS RESPONDED TO PRESSES APPROPRIATELY. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE KEYPAD BUTTONS HAVE NOT RESPONDED PROPERLY FOR THE PAST SEVERAL WEEKS. SHE CONFIRMED THAT THERE IS NO STRUCTURAL DAMAGE, BUT NOTED THAT THE KEYPAD SYMBOLS ARE WORN. THE PT DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS, AND STATED THAT SHE WEARS THE PUMP IN HER POCKET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020

Patients

Seq Age Sex Outcome Treatment
1 31 YR