FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL MEDIUM STRAIGHT ATTACHMENT
MDR report key: 2070824
·
Received April 15, 2011
Report
- Report Number
- 1811755-2011-01287
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 18, 2011
- Report Date
- March 18, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING FAILURE ANALYSIS, OVERHEATING WAS NOT DUPLICATED. HOWEVER, THE BEARING WAS DRY DEBRIS AND CORROSION WERE NOTED WITHIN THE DEVICE. THE DEBRIS AND CORROSION WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE STRYKER MICRO DRILL MEDIUM STRAIGHT ATTACHMENT IS AN IRREPARABLE PART ONCE IT IS DISASSEMBLED; AS A RESULT, THE DEVICE WAS NOT RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A ORAL MAX PROCEDURE. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE; ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO DRILL MEDIUM STRAIGHT ATTACHMENT | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |