FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 2070824 · Received April 15, 2011

Report

Report Number
1811755-2011-01287
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
March 18, 2011
Report Date
March 18, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, OVERHEATING WAS NOT DUPLICATED. HOWEVER, THE BEARING WAS DRY DEBRIS AND CORROSION WERE NOTED WITHIN THE DEVICE. THE DEBRIS AND CORROSION WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE STRYKER MICRO DRILL MEDIUM STRAIGHT ATTACHMENT IS AN IRREPARABLE PART ONCE IT IS DISASSEMBLED; AS A RESULT, THE DEVICE WAS NOT RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

A STRYKER MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A ORAL MAX PROCEDURE. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE; ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.