FDA Adverse Event
Malfunction
Summary report: N
TPS UNIVERSAL DRIVER
MDR report key: 2070821
·
Received April 15, 2011
Report
- Report Number
- 1811755-2011-01283
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 18, 2011
- Report Date
- March 25, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- K942956
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRIVER WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE RUNNING ON ITS OWN WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS CORROSION AND PROBLEMS WITH THE POTTED TRIGGER ASSEMBLY, DRIVESHAFT, BEARINGS, AND THE FLEX IN THE MOTOR ASSEMBLY. THOSE PARTS WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DRIVER CONTINUED TO RUN ON ITS OWN DURING A SMALL BONE FRACTURE REPAIR PROCEDURE. THE CASE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO MEDICAL INTERVENTION. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TPS UNIVERSAL DRIVER | HWE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |