FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2070690
·
Received April 15, 2011
Report
- Report Number
- 1811755-2011-01289
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 18, 2011
- Report Date
- March 18, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING TESTING. UPON FURTHER ANALYSIS, THE BEARINGS AND OTHER COMPONENT PARTS WERE DAMAGED. A DAMAGED SPINDLE HOUSING AND WORN BEARINGS WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE FAULTY PARTS WERE REPLACED, PREVENTIVE MAINTENANCE DONE AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE SAME DEVICE WITHOUT ANY PROCEDURE DELAY; NO ADVERSE EVENT WAS ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |