ARCHITECT TOTAL B-HCG
Report
- Report Number
- 1628664-2011-00185
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Report Date
- August 6, 2010
- Manufacturer
- A.I.D.D LONGFORD
- Product Code
- DHA
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE SUSPECT MEDICAL DEVICE HAS CHANGED FROM THE (B)(4). MFR # 3005094123-2011-00548 HAS BEEN SUBMITTED TO ADDRESS THIS ERROR
(B)(4). A PRODUCT DEFICIENCY HAS BEEN IDENTIFIED. WITH A UNIQUE COMBINATION OF INSTRUMENT AND REAGENT CONDITIONS THERE IS A POTENTIAL FOR SAMPLE CARRYOVER WHICH CAN RESULT IN FALSELY ELEVATED (B)(6) CONCENTRATIONS ON THE ARCHITECT I1000SR SYSTEM. NEGATIVE SAMPLES HAVE RETURNED (B)(6) AND GREY ZONE (B)(6) RESULTS WHEN A HIGH CONCENTRATION (B)(6) SAMPLE IS ASSESSED PRIOR TO THE NEGATIVE SAMPLE WITH ARCHITECT TOTAL (B)(6) (LIST NUMBER 7K78/6C21) ON AN I1000SR INSTRUMENT WHICH ALSO UTILIZES THE ARCHITECT (B)(6) (LIST NUMBER 6C17) ASSAY. FURTHER INVESTIGATION OF THIS QUALITY ISSUE WILL BE CONDUCTED. AN INVESTIGATION IS IN PROCESS.
THE CUSTOMER STATED AN ARCHITECT I1000SR ANALYZER GENERATED FALSELY ELEVATED (B)(6) RESULTS FOR THREE PATIENT SAMPLES. DATA WAS REVIEWED AND FOUND NOT TO AFFECT THE CLINICAL INTERPRETATION OR MEDICAL DECISION MAKING. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT TOTAL B-HCG | AUTOMATED IMMUNOASSAY ANALYZER | DHA | A.I.D.D LONGFORD | 81919JN00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCH BHCG REAGENT LN 7K78-25 LOT 81919JN00| ARCH I1000SR LN 01L86-01 (B)(4) |