OT ULTRALINK METER
Report
- Report Number
- 2939301-2011-03441
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Report Date
- April 7, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (05/20/2011)-DEVICE EVALUATION: THE LAY USER/PATIENT'S PRODUCT(S) INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: ON (B)(6), 2011, THE CONTROL SOLUTION WAS FOUND TO HAVE LOW GLUCOSE. ON (B)(6), 2011, THE TEST STRIPS WERE TESTED AND PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K073231.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE INACCURATE CONTROL LOW WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3080722 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |