HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-05084
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- September 19, 2010
- Report Date
- October 4, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SYSTEM ERROR 2240 (AIR IN LINE) ALARM WAS IDENTIFIED DURING A REVIEW OF A RETURNED HOMECHOICE DEVICE. THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION, THEREFORE, THE SAMPLE WAS NOT REQUESTED. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE; THEREFORE, THE CAUSE WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE SYSTEM ERROR OCCURRED ON (B)(6) 2010. A SE 2240 (AIR IN SET) IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | HOMECHOICE |