FDA Adverse Event Malfunction Summary report: N

ENDO GIA

MDR report key: 20706138 · Received November 18, 2024

Report

Report Number
2647580-2024-04832
Event Type
Malfunction
Date Received
November 18, 2024
Date of Event
October 18, 2024
Report Date
January 24, 2025
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDW
UDI-DI
10884523003215
PMA / PMN Number
K111825
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE RELOAD WAS FULLY FIRED. THE CLAMPING MECHANISM WAS DEFORMED. THE RELOAD HAD DAMAGE TO THE CUTTING EDGE OF THE KNIFE BLADE. IT WAS REPORTED THAT THE STAPLES DID NOT FORM PROPERLY. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THIS ISSUE MAY OCCUR UNDER THE FOLLOWING CONDITIONS. APPLICATION OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE. APPLICATION WITH AN OBSTACLE INCORPORATED IN THE JAWS. IN ANY OF THESE CIRCUMSTANCES, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY AND TISSUE MAY NOT BE FULLY TRANSECTED. THE DAMAGED KNIFE BLADE MAY OCCUR WHEN AN OBSTACLE HAS BEEN INCORPORATED IN THE JAWS DURING APPLICATION. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FAILURE TO COMPLETELY FIRE THE RELOAD WILL RESULT IN AN INCOMPLETE CUT AND/OR INCOMPLETE STAPLE FORMATION, WHICH MAY RESULT IN POOR HEMOSTASIS AND/OR LEAKAGE. ALWAYS INSPECT THE TISSUE THICKNESS AND SELECT AN APPROPRIATE STAPLE SIZE PRIOR TO APPLICATION OF THE ENDO GIA¿ ULTRA UNIVERSAL SHORT, ENDO GIA¿ ULTRA UNIVERSAL OR ENDO GIA¿ ULTRA UNIVERSAL XL STAPLER. OVERLY THICK OR THIN TISSUE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION. WHEN POSITIONING THE STAPLER ON THE APPLICATION SITE, ENSURE THAT NO OBSTRUCTIONS, SUCH AS CLIPS, ARE INCORPORATED INTO THE INSTRUMENT JAWS. FIRING OVER AN OBSTRUCTION MAY RESULT IN INCOMPLETE CUTTING ACTION AND/OR IMPROPERLY FORMED STAPLES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCTS: EGIA60AMT, EGIA60AMT EGIA 60 ARTIC MED THICK SULU (LOT#P2D0508S); EGIAUXL, EGIAUXL ENDOGIA ULTRA UNIV XL STAPLER (LOT#P2F0439S). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY PROCEDURE, WHILE CUTTING AND CLOSING THE GASTRIC ANTRUM AND STOMACH BODY, THE STAPLES WERE NOT FORMED PROPERLY, AND THE TISSUE OOZED BLOOD. THE SAME SITUATION OCCURRED WITH THE SECOND RELOAD. TO RESOLVE THE ISSUE, HAND-STITCHING WITH SUTURES WAS USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2112030 ENDO GIA STAPLE, IMPLANTABLE GDW US SURGICAL PUERTO RICO EGIA60AMT P2D0508S 10884523003215

Patients

Seq Age Sex Outcome Treatment
1 NA Female