ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN
Report
- Report Number
- 9610824-2011-00045
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- March 14, 2011
- Report Date
- April 12, 2011
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMBH
- Product Code
- LKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PHYSICIAN
Narratives
STN# 125098.
THE CUSTOMER COMPLAINED A FALSE NEGATIVE REACTION OF A PATIENT SAMPLE WITH ANTI-HUMAN GLOBULIN, ANTI-IGG SOLIDSCREEN II. THE SAMPLE WAS SUPPOSED TO CONTAIN AN ANTI-JK(B). WE RECEIVED THE PATIENT SAMPLE BUT NOT THE COMPLAINED LOT OF ANTI-HUMAN GLOBULIN, ANTI-IGG SOLIDSCREEN II. THE PATIENT SAMPLE WAS TESTED WITH THE RETENTION SAMPLE ON TANGO IN THE QUALITY CONTROL LABORATORY AND REACTED NEGATIVELY. FURTHERMORE, THE SAMPLE WAS TESTED ON GELCARD AND REACTED ALSO NEGATIVE. A FURTHER TESTING SHOWED A POSITIVE REACTION OF THE SAMPLE ONLY IN THE POLYETHYLENEGLYCOL (PEG) METHOD. THE CORRECT FUNCTION OF THE AFFECTED LOT ANTI-HUMAN GLOBULIN, ANTI-IGG SOLIDSCREEN II WAS CONFIRMED BY TESTING DIFFERENT WEAK ANTIBODIES. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN | ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN | LKJ | BIO-RAD MEDICAL DIAGNOSTICS GMBH | 7016110-05 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TANGO OPTIMO: SERIAL # (B)(4)| ORTHO IGG GEL CARD: LOT 102010001-23| ORTHO 0, 8% SCREENING CELLS: LOT VSS370| EXP. 09/02/2011| EXP. 04/05/2011| BIOTESTCELL 3: (816085100), LOT 8107011| MLB2: (805200100), LOT 7036120, EXP. 08/28/2012| ORTHO RESOLVE ANEL A: LOT VRA153, EXP. 04/05/2011| EXP. 04/05/2011| SOLIDSCREEN II STRIP: (806521100), LOT 7039040| EXP. 09/23/2012 |