DEROYAL 1500CC CANISTER KIT, 2 ELBOWS
Report
- Report Number
- 2320762-2011-00008
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- October 18, 2010
- Report Date
- October 21, 2010
- Manufacturer
- DEROYAL INDUSTRIES, INC.
- Product Code
- DWD
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
Narratives
THE INCIDENT DEVICE WAS NOT RETURNED AND NO LOT NUMBER WAS PROVIDED FOR THE INVESTIGATION OF THE REPORTED MALFUNCTION. THE MANUFACTURER HAS DETERMINED THAT POSSIBLE ROOT CAUSES MAY BE: TUBING INADVERTENTLY DISCONNECTED FROM THE ELBOW OR A POOR CONNECTION. THE LID COULD HAVE AN OCCLUDED PORT DUE TO MOLD DAMAGE WHICH WOULD BE A MANUFACTURING DEFECT. THE LID COULD HAVE A SHORT SHOT (NON-FILL) OF A CAP, WHICH WOULD ALLOW VACUUM TO ESCAPE FROM THE COLLECTION UNIT. THIS WOULD BE A MANUFACTURING DEFECT. THE LID COULD HAVE BEEN HOOKED UP INCORRECTLY WITH PATIENT TUBING PLACED ON THE VACUUM PORT. BECAUSE OUR INVESTIGATION FINDINGS ARE INCONCLUSIVE WITH THE INFORMATION PROVIDED AND THE POSSIBLE DEFECTS NOTED ABOVE WERE NOT FOUND IN THE INVENTORY INSPECTION CONDUCTED, NO FURTHER ACTION CAN BE TAKEN AT THIS TIME. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.
WHEN TRACHEAL SUCTIONING WAS ATTEMPTED IN A PATIENT'S ROOM, THE SUCTION WOULD NOT WORK. AGAIN ATTEMPTED SUCTION WITH A NEW LID AND CANISTER. TRIED ANOTHER SUCTION MACHINE, STILL DID NOT WORK. ANOTHER NURSE AND RESPIRATORY THERAPIST TRIED WITH THE SAME CANISTER AND LID AND STILL NO SUCTION. CODE BLUE WAS CALLED AND A SUCTION MACHINE AND CANISTER FROM THE CRASH CAST WAS USED WHICH DID SUCTION THE PATIENT. SUCTIONING WAS COMPLETED AND THE PATIENT WAS GIVEN EMOTIONAL SUPPORT AND REASSURANCE. HE WAS ASSURED THAT EXTRA SUCTION EQUIPMENT WOULD BE AVAILABLE IN HIS ROOM FOR BACK-UP. THE CANISTERS WERE DEFECTIVE AND WOULD NOT SUCTION. THE CANISTER WILL BE RETURNED TO DEROYAL INDUSTRIES, INC BY THE SPD DEPARTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEROYAL 1500CC CANISTER KIT, 2 ELBOWS | CANISTER, SUCTION | DWD | DEROYAL INDUSTRIES, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Life Threatening| R |