FDA Adverse Event Malfunction Summary report: N

DEROYAL 1500CC CANISTER KIT, 2 ELBOWS

MDR report key: 2070579 · Received April 13, 2011

Report

Report Number
2320762-2011-00008
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
October 18, 2010
Report Date
October 21, 2010
Manufacturer
DEROYAL INDUSTRIES, INC.
Product Code
DWD
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT DEVICE WAS NOT RETURNED AND NO LOT NUMBER WAS PROVIDED FOR THE INVESTIGATION OF THE REPORTED MALFUNCTION. THE MANUFACTURER HAS DETERMINED THAT POSSIBLE ROOT CAUSES MAY BE: TUBING INADVERTENTLY DISCONNECTED FROM THE ELBOW OR A POOR CONNECTION. THE LID COULD HAVE AN OCCLUDED PORT DUE TO MOLD DAMAGE WHICH WOULD BE A MANUFACTURING DEFECT. THE LID COULD HAVE A SHORT SHOT (NON-FILL) OF A CAP, WHICH WOULD ALLOW VACUUM TO ESCAPE FROM THE COLLECTION UNIT. THIS WOULD BE A MANUFACTURING DEFECT. THE LID COULD HAVE BEEN HOOKED UP INCORRECTLY WITH PATIENT TUBING PLACED ON THE VACUUM PORT. BECAUSE OUR INVESTIGATION FINDINGS ARE INCONCLUSIVE WITH THE INFORMATION PROVIDED AND THE POSSIBLE DEFECTS NOTED ABOVE WERE NOT FOUND IN THE INVENTORY INSPECTION CONDUCTED, NO FURTHER ACTION CAN BE TAKEN AT THIS TIME. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

WHEN TRACHEAL SUCTIONING WAS ATTEMPTED IN A PATIENT'S ROOM, THE SUCTION WOULD NOT WORK. AGAIN ATTEMPTED SUCTION WITH A NEW LID AND CANISTER. TRIED ANOTHER SUCTION MACHINE, STILL DID NOT WORK. ANOTHER NURSE AND RESPIRATORY THERAPIST TRIED WITH THE SAME CANISTER AND LID AND STILL NO SUCTION. CODE BLUE WAS CALLED AND A SUCTION MACHINE AND CANISTER FROM THE CRASH CAST WAS USED WHICH DID SUCTION THE PATIENT. SUCTIONING WAS COMPLETED AND THE PATIENT WAS GIVEN EMOTIONAL SUPPORT AND REASSURANCE. HE WAS ASSURED THAT EXTRA SUCTION EQUIPMENT WOULD BE AVAILABLE IN HIS ROOM FOR BACK-UP. THE CANISTERS WERE DEFECTIVE AND WOULD NOT SUCTION. THE CANISTER WILL BE RETURNED TO DEROYAL INDUSTRIES, INC BY THE SPD DEPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEROYAL 1500CC CANISTER KIT, 2 ELBOWS CANISTER, SUCTION DWD DEROYAL INDUSTRIES, INC.

Patients

Seq Age Sex Outcome Treatment
1 61 YR Life Threatening| R