FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM

MDR report key: 20705667 · Received November 18, 2024

Report

Report Number
0002023141-2024-03716
Event Type
Injury
Date Received
November 18, 2024
Date of Event
June 27, 2024
Report Date
November 17, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER: (B)(4). G4: PREMARKET IDENTIFICATION K061410/K011028/K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCE'S AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. H3 OTHER TEXT: SUMMARY INVESTIGATION.

Description of Event or Problem · 0

DOCTOR REPORTED THAT IMPLANT WITH CEMENTED CROWN AT TOOTH SITE 26 WAS REMOVED DUE TO PERI-IMPLANTITIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2037356 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL 61195992

Patients

Seq Age Sex Outcome Treatment