FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2070541 · Received March 29, 2011

Report

Report Number
2531779-2011-02028
Event Type
Malfunction
Date Received
March 29, 2011
Report Date
March 1, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. DURING INVESTIGATION, THE UP AND DOWN ARROW KEY CONTACTS WERE OBSERVED TO BE MISALIGNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DOWN ARROW REQUIRES MULTIPLE PRESSES FOR A RESPONSE. IT WAS REPORTED THAT THE RUBBER IS INTACT AND THE PUMP IS NOT EXPOSED TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1