FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2070539 · Received March 29, 2011

Report

Report Number
2531779-2011-02018
Event Type
Malfunction
Date Received
March 29, 2011
Report Date
February 28, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. THE UP ARROW AND CONTRAST KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE DURING TESTING. DURING INVESTIGATION, THE CONTRAST BUTTON WAS OBSERVED TO BE MISALIGNED. ALL KEYPAD BUTTONS DID NOT SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP ARROW IS RESPONDING POORLY TO PRESSES. IT WAS REPORTED THAT THE KEYPAD IS INTACT AND THE PUMP IS NOT EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1