FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2070524
·
Received March 29, 2011
Report
- Report Number
- 2531779-2011-02043
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Report Date
- February 25, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. A REVIEW OF PUMP HISTORY DID NOT FIND ANY EVIDENCE OF THE LOAD STEP STOPPING SHORT. PUMP WAS LOADED AND PRIMED SEVERAL TIMES WITHOUT DUPLICATING THE COMPLAINT. DURING EVAL THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.
Description of Event or Problem · 1
PUMP IS RETURNED FOR SHORT LOAD STEP CAUSING A LARGE PRIME VOLUME. EVAL REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVAL RESULTS.