FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2070524 · Received March 29, 2011

Report

Report Number
2531779-2011-02043
Event Type
Malfunction
Date Received
March 29, 2011
Report Date
February 25, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. A REVIEW OF PUMP HISTORY DID NOT FIND ANY EVIDENCE OF THE LOAD STEP STOPPING SHORT. PUMP WAS LOADED AND PRIMED SEVERAL TIMES WITHOUT DUPLICATING THE COMPLAINT. DURING EVAL THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.

Description of Event or Problem · 1

PUMP IS RETURNED FOR SHORT LOAD STEP CAUSING A LARGE PRIME VOLUME. EVAL REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVAL RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. IR 1250 NA

Patients

Seq Age Sex Outcome Treatment
1