FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2070376 · Received April 18, 2011

Report

Report Number
2032227-2011-00946
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THE REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE HAS BEEN EXPERIENCING HIGH BLOOD GLUCOSE LEVELS OVER 400 MG/DL DUE TO BENT CANNULAS. THE CUSTOMER ALSO STATED THAT THE INSULIN PUMP HAS NOT GIVEN HER A NO DELIVERY ALARM. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE CUSTOMER DIDN'T HAVE A TUBING CLAMP FOR THE HIGH PRESSURE TEST. A TUBING CLAMP WAS SENT TO THE CUSTOMER. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 31 YR