FDA Adverse Event Malfunction Summary report: N

GAMMAPLAN

MDR report key: 207034 · Received January 20, 1999

Report

Report Number
9612186-1999-00001
Event Type
Malfunction
Date Received
January 20, 1999
Date of Event
December 22, 1998
Report Date
December 22, 1998
Manufacturer
ELEKTA INSTRUMENT AB
Product Code
LMX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

WHILE USING THE TREATMENT PLANNING SOFTWARE, THE PHYSICIAN USED THE PLUG SHIELDS FEATURE, AND THEN MERGED THE PLUG PATTERNS. HE THEN CREATED A NEW PLAN AND CHOSE THE "COPY EXISTING PLAN" FEATURE. UPON COPYING, THE LEKSELL GAMMAPLAN SOFTWARE GAVE AN ERROR STATING THAT PLUG PATTERN "HELMET 18A NOT FOUND." THE SHOT MENU SHOWED ONLY THE PATTERN "NONE" FOR EACH OF THE SHOTS. THE PLUG MENU SHOWED NO PLUGS. HOWEVER, THE PROTOCOL PRINTOUT GAVE TIMES AS IF THE PLUGS STILL EXISTED WITHOUT PRINTING ANY PLUGS AND THE ISODOSE CURVES BEHAVED AS IF THE PLUGS WERE STILL THERE. THIS WAS TRUE FOR BOTH THE ORIGINAL PLAN AND FOR THE COPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GAMMAPLAN TREATMENT PLANNING SOFTWARE LMX ELEKTA INSTRUMENT AB VERSION 5.12 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other