FDA Adverse Event
Malfunction
Summary report: N
GAMMAPLAN
MDR report key: 207034
·
Received January 20, 1999
Report
- Report Number
- 9612186-1999-00001
- Event Type
- Malfunction
- Date Received
- January 20, 1999
- Date of Event
- December 22, 1998
- Report Date
- December 22, 1998
- Manufacturer
- ELEKTA INSTRUMENT AB
- Product Code
- LMX
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
WHILE USING THE TREATMENT PLANNING SOFTWARE, THE PHYSICIAN USED THE PLUG SHIELDS FEATURE, AND THEN MERGED THE PLUG PATTERNS. HE THEN CREATED A NEW PLAN AND CHOSE THE "COPY EXISTING PLAN" FEATURE. UPON COPYING, THE LEKSELL GAMMAPLAN SOFTWARE GAVE AN ERROR STATING THAT PLUG PATTERN "HELMET 18A NOT FOUND." THE SHOT MENU SHOWED ONLY THE PATTERN "NONE" FOR EACH OF THE SHOTS. THE PLUG MENU SHOWED NO PLUGS. HOWEVER, THE PROTOCOL PRINTOUT GAVE TIMES AS IF THE PLUGS STILL EXISTED WITHOUT PRINTING ANY PLUGS AND THE ISODOSE CURVES BEHAVED AS IF THE PLUGS WERE STILL THERE. THIS WAS TRUE FOR BOTH THE ORIGINAL PLAN AND FOR THE COPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GAMMAPLAN | TREATMENT PLANNING SOFTWARE | LMX | ELEKTA INSTRUMENT AB | VERSION 5.12 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |