FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2070269 · Received April 19, 2011

Report

Report Number
3004209178-2011-81104
Event Type
Injury
Date Received
April 19, 2011
Date of Event
April 7, 2011
Report Date
April 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

UNABLE TO PRIME DURING THE PRIME TEST DUE TO THE FAULTY FORCE SENSOR. UNABLE TO PERFORM THE BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY TEST DUE TO THE PRIME/FILL ANOMALY. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 439 MG/DL AND SYMPTOMS OF DIZZINESS, NAUSEA, VOMITING, AND EXCESSIVE URINATION. THE CUSTOMER STATED THAT SHE USES A QUICK-SET INFUSION SET AND THAT SHE SOMETIMES HAS BENT CANNULA. PROGRAMMING WAS CORRECT. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP PASSED BOTH THE FIXED PRIME AND THE HIGH PRESSURE TESTS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization UNOMEDICAL QUICK-SET INSULIN INFUSION SET,| LOT 9202038