PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2011-81104
- Event Type
- Injury
- Date Received
- April 19, 2011
- Date of Event
- April 7, 2011
- Report Date
- April 14, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- NOT APPLICABLE
Narratives
UNABLE TO PRIME DURING THE PRIME TEST DUE TO THE FAULTY FORCE SENSOR. UNABLE TO PERFORM THE BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY TEST DUE TO THE PRIME/FILL ANOMALY. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 439 MG/DL AND SYMPTOMS OF DIZZINESS, NAUSEA, VOMITING, AND EXCESSIVE URINATION. THE CUSTOMER STATED THAT SHE USES A QUICK-SET INFUSION SET AND THAT SHE SOMETIMES HAS BENT CANNULA. PROGRAMMING WAS CORRECT. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP PASSED BOTH THE FIXED PRIME AND THE HIGH PRESSURE TESTS. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization | UNOMEDICAL QUICK-SET INSULIN INFUSION SET,| LOT 9202038 |