FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2070183
·
Received April 27, 2011
Report
- Report Number
- 2531779-2011-02910
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 1, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE MANUFACTURE DATE: 10/2009. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. INVESTIGATION REVEALED THAT THE AUDIO BOLUS BUTTON COVER IS PEELING. EVALUATION CONFIRMED THAT THE AUDIO BOLUS BUTTON IS DIFFICULT TO PRESS. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER THE AUDIO BOLUS BUTTON KEY CONTACT. TESTING REVEALED THAT ALL KEYPAD BUTTONS ARE RESPONDING APPROPRIATELY.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE AUDIO BOLUS BUTTON REQUIRES EXCESSIVE FORCE TO OBTAIN A RESPONSE. HE NOTED THAT THE ISSUE BEGAN 2 WEEKS AGO, AND THE AUDIO BOLUS BUTTON COVER CAME OFF TODAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |