ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-02908
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 1, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4):THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. INSPECTION REVEALED THAT THE KEYPAD IS PEELING AT THE OK KEYPAD BUTTON. EVALUATION CONFIRMED THAT THE DOWN ARROW KEYPAD BUTTON IS INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE DOWN ARROW KEYPAD BUTTON IS STICKING AND REQUIRES MULTIPLE PRESSES TO OBTAIN A RESPONSE. SHE NOTED THAT THE DOWN ARROW BUTTON CLICKS INTERMITTENTLY WHEN PRESSED. THE PATIENT STATED THAT THE DOWN ARROW BUTTON BECAME INTERMITTENTLY UNRESPONSIVE A WEEK AGO, AND TODAY THE KEYPAD BEGAN PEELING. SHE CONFIRMED THAT ALL OTHER KEYPAD BUTTONS ARE RESPONDING APPROPRIATELY AND CLICK BACK AS EXPECTED. THE PATIENT DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |