FDA Adverse Event Malfunction Summary report: N

PUMP,COLLEAGUE SINGLE CHANNELCOLOUR

MDR report key: 2070078 · Received April 26, 2011

Report

Report Number
6000001-2011-03203
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
September 3, 2010
Report Date
November 15, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED CONDITION OF BATTERY DEPLETED SET ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY. THE ROOT CAUSE HAS NOT BEEN DETERMINED AT THIS TIME. NO REPAIRS HAVE BEEN PERFORMED AT THIS TIME SINCE THIS IS A BAXTER OWNED DEVICE. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Additional Manufacturer Narrative · 1

THERE IS AN ONGOING CAPA INVESTIGATION, (B)(4) ASSOCIATED WITH THIS REPORT. (B)(4).

Description of Event or Problem · 1

COLLEAGUE INFUSION PUMP WAS REPORTED ORIGINALLY FOR A BATTERY ISSUE. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, THE DEVICE WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET ALARM WHICH INTERRUPTED DELIVERY. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.92 CATEGORIZED AS REMEDIATED.