FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2070074 · Received April 26, 2011

Report

Report Number
2531779-2011-02902
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 28, 2011
Report Date
March 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER THE EVALUATION IS COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. KEYPAD BUTTON PRESSES DID ACTIVATE DESIRED PUMP FUNCTIONS. THE KEYPAD WAS REMOVED AND NO DEFECTS WERE FOUND.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 68 YR