FDA Adverse Event Malfunction Summary report: N

SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER

MDR report key: 2070068 · Received April 26, 2011

Report

Report Number
2050012-2011-01301
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 27, 2011
Report Date
March 27, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K188495
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER ATTEMPTED TO TROUBLESHOOT BY SPLICING THE TUBING TOGETHER. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND REPLACED THE REAGENT SIDE WASH STATION WASTE LINE AND NO FURTHER ISSUES OCCURRED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT A TUBE BROKE IN THE HYDRO AREA OF THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER, CAUSING FLUID TO SPILL OUT. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. CX5 DELTA N/A